Description
Ph Mode:Up to 5 points push-button calibration with auto-buffer recognition. Selectable pH buffer standards, (USA, DIN, NIST, User-defined) o o resolution (0.1, 0.01, 0.001pH )temperature units ( C or F) Automatically calculates and displays electrode slope that helps you decide whether to replace pH electrode.Calibration due reminder prompts you to calibrate the meter regularly.
ORP Mode:Absolute or Relative mV measurements provides accurate ORP measured value .1 point offset calibration allows user adjusts displayed value to a known standard.
CONDUCTIVITY/TDS/SALINITY/RESISTIVITY MODE:: Up to 5-point push-button calibration allows user to uses customizedcalibration solutions. Default 10uS, 84uS, 1413uS, 12.88mS and 111.8mS)Adjustable TDS conversion factors, cell constants, temperature coefficients and o o normalization temperature. (20 C or 25 C)Allows user selects linear or pure water temperature compensation modes.In salinity measurement modes, selectable practical salinity (PSU) or seawater measurement standards. (ppt)
| SPECIFICATION | |
|---|---|
| Parameter | Labman Multiparameter |
| Model | LMMP30 |
| pH Range | |
| pH Range | -2.000 ~ 20.000 pH |
| Resolution | 0.001, 0.01, 0.1 pH, Selectable |
| Accuracy | ±0.002 pH |
| Calibration Points | Up to 5 points |
| pH Buffer Option |
USA (pH 1.68, 4.01, 7.00, 10.01, 12.45), NIST (pH 1.68, 4.01, 6.86, 9.18, 12.45), DIN (pH 1.09, 4.65, 6.79, 9.23, 12.75) |
| ORP | |
| ORP (mV) Range / Calibration | 0 ~ ±1999.9 mV / 1 point |
| Resolution | 0.1 / 1 mV, Selectable |
| Accuracy | ±0.2 mV |
| Measuring Mode | Absolute mV or Relative mV |
| Conductivity | |
| Conductivity Range | 0 ~ 200 mS/cm |
| Resolution | 0.01 / 0.1 / 1 |
| Accuracy | ±0.5% Full Scale |
| Cell Constant | 0.1 / 1 / 10 / User defined |
| Temperature Coefficient | 0.0 ~ 10.0% per °C |
| Calibration Solutions | 10 µS/cm, 84 µS/cm, 1413 µS/cm, 12.88 mS/cm, 111.8 mS/cm |
| Compensation Modes | Linear or Pure Water |
| TDS | |
| TDS Range | 0 ~ 10 ppt (Max. 200 ppt, depending on factor setting) |
| Resolution | 0.01 / 0.1 / 1 |
| Accuracy | ±1% Full Scale |
| TDS Factor | 0.0 ~ 1.0 |
| Salinity | |
| Salinity Range | 0 ~ 80 ppt |
| Resolution | 0.01 / 0.1 / 1 |
| Accuracy | ±1% Full Scale |
| Measurement Mode | Practical Salinity (PSU) or Seawater Measurement Standard (ppt) |
| Resistivity | |
| Resistivity Range | 0 ~ 100 MΩ |
| Resolution | 1 / 0.1 / 0.01 |
| Accuracy | ±1% Full Scale |
| Temperature | |
| Temperature Range | 0 ~ 100°C / 32 ~ 212°F |
| Temperature Compensation | Automatic or Manual |
| Temperature Accuracy | ±0.5°C / ±0.9°F |
| General | |
| Auto-Off Function | 10 / 20 / 30 minutes after last key pressed |
| Reset & Auto Hold Function | Yes |
| Electrode Diagnosis | Slope, Offset |
| Stability Conditions | High or Low, Selectable |
| Calibration Due Reminder | 1 to 30 days, Selectable |
| Memory | 500 data sets with Date/Time |
| Connector | BNC 6-pin connector |