Maximum Capacity[Max] |
6.1 g |
6 / 21 g |
Minimum load |
0.1 mg |
0.1 mg |
Readability [d] |
1 µg |
1/2 µg |
Tare range |
-6.1 g |
-21 g |
Standard repeatability [5% Max] |
0.8 µg |
1.3 µg |
Standard repeatability [Max] |
2.5 µg |
3.5 µg |
Standard minimum weight (USP) |
1.6 mg |
2.6 mg |
Standard minimum weight (U=1%, k=2) |
0.16 mg |
0.26 mg |
Permissible repeatability [5% Max] |
1.5 µg |
2 µg |
Permissible repeatability [Max] |
3 µg |
5 µg |
Linearity |
±7 µg |
±9 µg |
Eccentric load deviation |
7 µg |
15 µg |
Stabilization time |
max 4s |
max 4s |
Adjustment(Calibration) |
internal (automatic) |
internal (automatic) |
OIML Class |
I |
I |
Display |
5.7” resistive colour touchscreen |
5.7” resistive colour touchscreen |
Communication interface |
2×RS232, 2×USB-A, Ethernet, 4 IN / 4 OUT (digital), Wi-Fi® |
2×RS232, 2×USB-A, Ethernet, 4 IN / 4 OUT (digital), Wi-Fi® |
Power supply |
100 ÷ 240 V AC 50 / 60 Hz |
100 ÷ 240 V AC 50 / 60 Hz |
Power consumption max. |
10 W |
10 W |
Operating temperature |
+10 ÷ +40 °C |
+10 ÷ +40 °C |
Operating temperature change rate |
±0.3°C/1h (±1°C/8h) |
±0.3°C/1h (±1°C/8h) |
Relative humidity |
40% ÷ 80% |
40% ÷ 80% |
Relative humidity change rate |
±1%/h (±4%/8h) |
±1%/h (±4%/8h) |
Weighing chamber dimensions |
ø168×163x228 mm |
ø199×170x217 mm |
Weighing pan dimensions |
ø30 mm |
ø30 mm |
Packaging dimensions |
720×385×485 mm |
720×385×485 mm |
Net Weight |
9.8 kg |
14.5 kg |
Gross weight |
14.3 kg |
18.9 kg |